Dileep Kumar Shetty and Ashapurna Marndi
Anisotropic Magnetoresistive (AMR) magnetometers are widely used in precision magnetic-field measurements; however, their measurement accuracy can be affected by temperature-dependent sensor characteristics, nonlinearities, cross-axis coupling, and calibration uncertainties. This study presents a hybrid calibration framework that integrates physics-based modelling with neural network-based residual correction and uncertainty quantification for temperature-aware AMR magnetometers. The proposed approach first employs a physics-based calibration model to characterize and compensate for temperature-dependent scale-factor variations, sensor misalignment, magnetic offsets, and cross-axis coupling. A neural network is subsequently used to learn and correct the residual nonlinear errors that remain after physics-based calibration. In addition, an uncertainty quantification framework is incorporated to assess the reliability of the calibrated magnetic-field measurements. The proposed methodology is evaluated using experimental AMR magnetometer data over varying magnetic-field and temperature conditions. The results demonstrate substantial improvements in calibration accuracy and measurement reliability compared with uncalibrated measurements, highlighting the potential of the proposed hybrid framework for accurate and uncertainty-aware magnetic-field metrology.